Table 3: Annual QC tests and tolerances levels for X-ray CT.

Test quantity

Acceptable

Achievable

CT alignment lights

± 5 mm

± 1 mm

SPR accuracy

± 2 mm

± 1 mm

CT number

± 5 from baseline valuea

± 4

Image Noise

± 25% from baseline valuea

±10% of the baseline

Uniformity

± 10

± 4

Artefact

No artefacts to affect diagnostic confidence

No visible artefacts

CTDIvol

< ± 20% manufacturer’s recommendations and console displayed dose values