Table 2: Average absorbed dose in some organ positions in the phantom during head CT scans with and without bismuth shielding on eyes.

Organ position Average absorbed dose (mGy)
Without bismuth shielding With bismuth shielding Head tilted
Lenses 17.44 ± 0.66a 9.49 ± 0.57 2.46 ± 0.57
Thyroid 2.35 ± 0.30 1.05 ± 0.41 1.33 ± 0.45
Pharynx 2.46 ± 0.28 1.20 ± 0.32 0.85 ± 0.29
Parotid Gland 22.97 ± 0.69 20.18 ± 0.81 3.90 ± 0.64
Salivary Gland 4.29 ± 0.55 2.65 ± 0.38 1.25 ± 0.32
Hypophysis 15.35 ± 0.47 13.79 ± 0.46 10.06 ± 0.67
Breast 0.94 ± 0.26 0.49 ± 0.22 0.58 ± 0.36
Spinal Cord 1.74 ± 0.56 0.48 ± 0.58 0.85 ± 0.38

aStandard deviation.