Table1: Patients' data, output of x-ray machine, value of exposure parameters as average (min-max) in the hospitals.

Hospital Name

Sample Size

Patients’ Data

Output of X-ray Machine

(mGy/mAs)

Exposure Parameters

Age of patient

(year)

Patient Thickness(tp) in cm

Kvp (kv)

mAs

FSD (cm) = FFD (cm) - tp (cm)

Public hospitals

H1

24

34 (17-55)

20.4 (16-30)

0.1204

75.9 (70-80)

47.0 (36-62.5)

84.6 (75-89)

H2

51

39 (18-74)

16.5 (11-30)

0.1022

74.6 (66-97)

32.4 (24-48)

88.5 (75-94)

Private hospitals

H3

42

38 (19-84)

22.4 (11-28)

0.0966

90.3 (83-98)

87.1 (60-120)

82.6 (77-94)

H4

28

37 (18-59)

16.9 (10-25)

0.0352

76.5 (70-85)

64.7 (50-80)

88.1 (80-95)