Table 3: Routine and optimized protocols.

Phantom

Protocol

Voltage

(kV)

Charge

(mA.s)

CTDIVOL

(mGy)

Adult

(16 cm)

Routine

120

200

39.92 ± 0.21*

Opt. 1

120

160

31.93 ± 0.16

Opt. 2

100

240

31.22 ± 0.28

Opt. 3

80

420

28.87 ± 0.18

Pediatric

(12 cm)

Routine

120

200

49.67 ± 0.82

Opt. 4

120

100

24.83 ± 0.41

Opt. 5

100

144

23.23 ± 0.18

Opt. 6

80

240

20.89 ± 0.20

*Standard deviation